Accuracy and Reoperation Rate for Cortical Bone Trajectory Screws
Accuracy and Reoperation Rate for Cortical Bone Trajectory Screws Utilizing the Cannulator Technique
Introduction
There has been an increased utilization of Cortical Bone Trajectory (CBT) screws in treatment of Lumbar Degenerative Spinal pathology. The benefits of this trajectory include easier ergonomics for screw placement and the ability to capture the middle column of the spine utilizing a minimally invasive single incision midline exposure. One of the challenges to widespread adoption of this technique include the novel “inside-out” trajectory and starting point identification which can be daunting for many surgeons. Given the cortical nature of the bone trajectory, many early techniques were clumsy utilizing multiple passes with burrs, hand drills, awls, and taps. The cannulator technique combines these instruments into one instrument and utilizes a slightly more lateral starting point to allow for more reproducible screw placement with minimal steps.
Materials/Methodology
A retrospective chart review was performed to identify all posterior lumbar fusion patients from January 1, 2014 through July 27, 2020. Once identified, postoperative plain radiographs were reviewed to obtain an accurate number of total posterior lumbar screws placed. Operative and office visit notes were then reviewed to determine the number of screws requiring repositioning after surgery. A screw revision rate was determined by dividing the total number of screws repositioned by the total number of screws placed.
Results
Between January 1, 2014 and July 27th, 2020, 1503 total posterior lumbar screws were placed. 5 patients returned to surgery for repositioning. In one case, the bilateral L4 screws were repositioned. 2 patients required a repositioned left L5 screw. One patient required a repositioned left L4 screw, and one required the right L4 screw. All screws repositioned were due to medial breaches. Only symptomatic patients received postoperative CT scans.
Discussion
There have been many previous studies looking at the placement of freehand standard trajectory pedicle screws compared with robotic navigated placement. This is the first paper that has looked at revision rates for malpositioned cortical screws. The modifications to both the technique and instrumentation are key to producing an extremely low revision/reoperation rate. The main weakness of this study is the lack of routine CT scans to determine the exact positioning of each cortical screw placed.